Abstract
A previously described vapour-pressure osmometer has been adapted for the measurement of depression of water potential (DPD) in samples of leaf tissue and soil. A pile of about 30 leaf annuli are contained in an air-tight metal capsule through the lid of which passes a pipette. A drop of water hangs from the tip of the pipette into the central cavity of the pile of leaf annuli. The drop is drawn bath into the pipette from time to time and the rate of evaporation from it measured directly. The water-potential depression is linearly related to the rate of evaporation. The instrument is calibrated empincally. A period of a hours is necessary to attain vapour equilibnum before readings are taken The readings then occupy less than 30 minutes. For the calibration data, using filter paper moistened with NaCl solution, the 5 percent inverse tolerance limits for single readings of water-potential depression were ±0.26 atm. With leaf material the errors appear to be about the same, although release of cell sap at the cut edges of the disks may cause an overestimation by as much as 0.5 atm. with turgid leaf material. Lining the capsule with leaf strips gave erroneous results. Using a suitably modified capsule the instrument can be used for soil-water potential measurement. Equilibration takes only 30 minutes so that a reading can be completed within an hour.
Abstract from the paper’s OpenAlex record, shown for indexing and research. Copyright remains with the authors or the publisher. Read at the source ↗
The text the model read, with the source record’s inline markup (italics, sub- and superscripts) rendered and its character entities decoded; quote offsets refer to the raw text.
Claims 6
Claim · model-written summary
Water-potential depression was linearly related to drop evaporation rate.
- Outcome
- water-potential depressione0
Not reportedTaxonComparatorDirectionValue
Verbatim
e0“The water-potential depression is linearly related to the rate of evaporation”
Claim · model-written summary
Single-reading tolerance limits were ±0.26 atm in filter-paper calibration.
- Outcome
- inverse tolerance limits for single readings of water-potential depressione0
- Value
- ±0.26 atme0
- Qualifier
- 5 percente0
Not reportedTaxon
Verbatim
e0“the 5 percent inverse tolerance limits for single readings of water-potential depression were ±0.26 atm”
Claim · model-written summary
Leaf-material errors appeared similar to calibration errors.
Not reportedTaxon
Verbatim
e0“With leaf material the errors appear to be about the same, although release of cell sap at the cut edges of the disks may cause an overestimation by as much as 0.5 atm”
Claim · model-written summary
Sap release at cut leaf-disk edges may overestimate water-potential depression by as much as 0.5 atm.
Not reportedTaxon
Verbatim
e0“release of cell sap at the cut edges of the disks may cause an overestimation by as much as 0.5 atm”
Claim · model-written summary
Leaf-strip lining yielded erroneous results.
Not reportedTaxonComparatorValue
Verbatim
e0“Lining the capsule with leaf strips gave erroneous results”
Claim · model-written summary
A modified capsule allowed soil-water potential readings within an hour after 30 minutes of equilibration.
Not reportedTaxon
Verbatim
e0“Using a suitably modified capsule the instrument can be used for soil-water potential measurement. Equilibration takes only 30 minutes so that a reading can be completed within an hour”